[Optical quality of three trifocal intraocular lens models : An optical bench comparison]
R Khoramnia1, T M Yildirim2, T Tandogan2
1David J Apple International Laboratory for Ocular Pathology and International Vision Correction Research Centre (IVCRC), Universitäts-Augenklinik Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Deutschland. Ramin.Khoramnia@med.uni-heidelberg.de.
Background:
Knowledge of the optical quality of different trifocal intraocular lenses (IOL) is important in customized patient care.
Objective:
Different trifocal IOL were compared regarding their optical quality.
Materials And Methods:
We analyzed the FineVision (PhysIOL, Liège, Belgium), the AT LISA tri 839MP (Zeiss, Oberkochen, Germany), and the AcrySofIQ PanOptix (Alcon, Fort Worth, TX, USA) with a power of +21D for the distance using the OptiSpheric IOL PRO optical bench (Trioptics, Wedel, Germany). The additions for the near and intermediate distances were as follows: +3.5D/+1.75D (FineVision), +3.33D/+1.66D (AT LISA tri), and + 3,25D/+ 2,17D (PanOptix). We evaluated the modulation transfer function (MTF) at a spatial frequency of 50lp/mm and the Strehl ratio using 3‑ (photopic) and 4.5-mm (mesopic) apertures.
Results:
The MTF at 50 lp/mm (FineVision/AT Lisa tri/PanOptix) at the far focus was 0.373/0.399/0.400 (3-mm aperture) and 0.512/0.311/0.243 (4.5-mm aperture). At the intermediate focus, the MTF was 0.162/0.147/0.153 (3-mm aperture) and 0.092/0.125/0.137 (4.5-mm aperture). The MTF at the near focus was 0.229/0.192/0.404 (3-mm aperture) and 0.217/0.212/0.169 (4.5-mm aperture). The Strehl ratio was 0.335/0.298/0.370 (3-mm aperture) and 0.243/0.180/0.270 (4.5-mm aperture) at the far focus. At intermediate distances, the Strehl ratio was 0.189/0.185/0.162 (3-mm aperture) and 0.099/0.097/0.114 (4.5-mm aperture). The Strehl ratio was 0.305/0.283/0.464 (3-mm aperture) and 0.177/0.181/0.155 (4.5-mm aperture) at the near focus.
Conclusion:
Evaluation of the three trifocal IOL models at the optical bench could show distinct peaks at the far, intermediate, and near focus. The results were comparable in terms of optical performance.