Imaging the Infant Chest without Sedation: Feasibility of Using Single Axial Rotation with 16-cm Wide-Detector CT
Yanan Zhu1, Zhengjun Li1, Jiankang Ma1
1From the Medical Imaging Centre (Y.Z., Z.L., J.M., Z.P., X.Q., M.X., H.Z.) and Department of Pediatrics (Y.H.), Ankang Central Hospital, Ankang 725000, China; and CT Laboratory, GE Healthcare China, Beijing, China (J.L.).
Insights
New computed tomography (CT) imaging for infant chests without sedation significantly reduces radiation dose and preparation time. This wide-detector CT method offers comparable image quality to traditional sedation-based scans.
Area of Science:
- Radiology
- Medical Imaging
- Pediatric Imaging
Background:
- Sedation in infants for computed tomography (CT) scans poses risks and increases preparation time.
- Conventional helical CT with sedation is a standard but time-consuming and radiation-intensive procedure for infant chest imaging.
Purpose of the Study:
- To compare image quality, patient preparation time, and radiation dose between single axial rotation 16-cm wide-detector CT without sedation and 64-row CT with sedation for infant chest imaging.
Main Methods:
- Prospective enrollment of 32 infants (group 1) for nonenhanced chest CT without sedation using a 16-cm wide-detector CT scanner.
- Comparison with a historical control group (group 2) of 30 infants who underwent 64-row CT with sedation.
- Utilized automatic tube current modulation and specific tube voltages based on patient weight.
Main Results:
- No significant differences in image quality (attenuation, noise, subjective scores) between the two groups.
- Group 1 showed significantly reduced scan time (83%), preparation time (57%), CT dose index (CTDI) (42%), and dose-length product (DLP) (52%) compared to group 2.
- Statistical significance was achieved for all reported reductions (P < .05).
Conclusions:
- Single axial rotation with 16-cm wide-detector CT without sedation is a viable alternative for infant chest imaging.
- This method effectively reduces radiation dose, preparation time, and scan duration.
- Comparable image quality is maintained, making it a potentially safer and more efficient option for pediatric patients.
Abstract:
Purpose To compare image quality, patient preparation time, and radiation dose using a single axial rotation with 16-cm wide-detector computed tomography (CT) in imaging the infant chest without sedation with those in infants examined by using a 64-row CT and sedation. Materials and Methods Thirty-two infants (group 1) were prospectively enrolled to undergo nonenhanced chest CT without sedation using a single axial rotation on a 16-cm wide-detector CT scanner. Patients were imaged with automatic tube current modulation and tube voltages of 80 kVp for patients weighing 5 kg or less and 100 kVp for patients weighing more than 5 kg. Patient preparation time, CT dose index (CTDI), dose-length product (DLP), and image quality were compared with those in a historical control group consisting of 30 infants (group 2) who underwent conventional helical scanning with sedation performed by using a 64-row volume CT scanner. The Student t test for independent samples was used to assess continuous variables. The Mann-Whitney rank test and the κ test were used to evaluate image quality. Results There was no statistically significant difference in body weight, age, mean CT attenuation value, image noise, and subjective image quality score between the two groups. However, compared with the group scanned by using a 64-row volume CT scanner (group 2), group 1 experienced significantly reduced scan time by 83% (0.35 second vs 2.01 seconds ± 0.21 [standard deviation]), preparation time by 57% (41.25 minutes ± 103.78 vs 96.5 minutes ± 151.77), CTDI by 42% (2.03 mGy ± 0.4 vs 3.52 mGy ± 0.03), and DLP by 52% (27.07 mGy·cm ± 6.97 vs 55.84 mGy·cm ± 6.46) (P < .05 for all). Conclusion Compared with conventional 64-row helical CT with sedation, use of a single axial rotation with 16-cm wide-detector CT in imaging the infant chest without sedation can reduce radiation dose, preparation time, and total scan time, while providing comparable image quality. © RSNA, 2017.
More Related Videos
02:09Multi-modal Pulmonary Imaging: Using Complementary Information from CT and Hyperpolarized 129Xe MRI to Evaluate Lung Structure-Function
Published on: April 12, 2024
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography