Electronic recording of transfusion-related patient observations: a comparison of two bedside systems
S Staples1, S Noel1, P Watkinson1
1Oxford NIHR Biomedical Research Centre, Oxford University Hospitals, University of Oxford, Oxford, UK.
Background And Objectives:
Vital sign observations should be monitored before, during and after transfusion to enable adverse events to be identified, but surveys in the UK show poor compliance with good practice. At the Oxford University Hospitals, there are two electronic bedside processes for recording observations; BloodTrack Tx (Haemonetics Corp.), the routine electronic transfusion process and a locally developed process, the System for Electronic Nursing Documentation (SEND) with integrated 'track and trigger' calculation for monitoring vital signs. The purpose of this study was to evaluate the conduct of patient observation monitoring for blood transfusion using two electronic bedside processes.
Materials And Methods:
This study examined the observations recorded during 200 single red cell unit transfusions.
Results:
186/200 (93%) transfusions had pretransfusion observations recorded using BloodTrack Tx. Mid-transfusion checks were performed during 133/200 (67%) of transfusions, of these checks most (87/200 (44%)) were documented as 'no apparent change' in observations. End transfusion observations were performed using BloodTrack Tx in 178/200 (89%). Both systems were frequently used, and staff had a preference for using SEND first for documenting the pretransfusion observations (102/116 (88%)) and at the end of a transfusion (75/115 (65%)).
Conclusion:
Electronic bedside systems result in improved monitoring of transfusion-related observations compared to manual processes based on data from UK surveys. There is increasing use of electronic systems in clinical practice; linkage between these two systems would prevent wasteful duplication of observations and could provide improved early warning of adverse events to transfusion compared to manual processes.
Related Concept Videos
Methods of Documentation VII: EMR
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Errors occurring during blood pressure monitoring
Several factors...
Methods of Documentation V: CBE
In CBE, healthcare professionals establish predefined standards of practice that define what constitutes...
Assessing Blood pressure in the Leg
Preparation:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...