A comparison study of stem taper material loss at similar and mixed metal head-neck taper junctions
D J Langton1, R P Sidaginamale2, T J Joyce2
1Freeman Hospital, Northern Retrieval Registry, University Hospital of North Tees and Freeman Hospital, Newcastle Upon Tyne, UK.
Aims:
We sought to determine whether cobalt-chromium alloy (CoCr) femoral stem tapers (trunnions) wear more than titanium (Ti) alloy stem tapers (trunnions) when used in a large diameter (LD) metal-on-metal (MoM) hip arthroplasty system.
Patients And Methods:
We performed explant analysis using validated methodology to determine the volumetric material loss at the taper surfaces of explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy (n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs with a rough male taper surface and a nominal included angle close to 5.666° were included. Multiple regression modelling was undertaken using taper angle, taper roughness, bearing diameter (horizontal lever arm) as independent variables. Material loss was mapped using a coordinate measuring machine, profilometry and scanning electron microscopy.
Results:
After adjustment for other factors, CoCr stem tapers were found to have significantly greater volumetric material loss than the equivalent Ti stem tapers.
Conclusion:
When taper junction damage is identified during revision of a LD MoM hip, it should be suspected that a male taper composed of a standard CoCr alloy has sustained significant changes to the taper cone geometry which are likely to be more extensive than those affecting a Ti alloy stem. Cite this article: Bone Joint J 2017;99-B:1304-12.
More Related Videos
10:49Intravital Longitudinal Imaging of Vascular Dynamics in the Calvarial Bone Marrow
Published on: April 11, 2025
12:18Co-localizing Kelvin Probe Force Microscopy with Other Microscopies and Spectroscopies: Selected Applications in Corrosion Characterization of Alloys
Published on: June 27, 2022
Related Concept Videos
Residual Stresses in Circular Shafts
Thin-Walled Hollow Shafts
Mechanical Characteristics of Steel
The tension test is fundamental for determining tensile strength. In this test, a steel specimen is stretched using a gripping device until it breaks. The data collected during this test are used...
Design of Transmission Shafts - Stress Analysis
Stress Concentrations in Circular Shafts
Yield Criteria for Ductile Materials under Plane Stress
The Maximum Shearing Stress Criterion, also known as...