Validity and reliability of ankle dorsiflexion measures in children with cerebral palsy
Do-Hyun Kim1, Duk-Hyun An2, Won-Gyu Yoo2
1HIP and MAL Laboratory, Department of Rehabilitation Science, Inje University, Gimhae, Korea.
Purpose:
We compared a goniometer method in a non-weight-bearing position with a tape measure method in a weight-bearing position to determine which was more reliable for assessing dorsiflexion range of motion (ROM) in children with cerebral palsy (CP).
Methods:
Ankle dorsiflexion ROM was measured using goniometer and tape measure methods in non-weight- and weight-bearing positions, respectively.
Results:
In the test-retest reliability of ankle dorsiflexion ROM using a universal goniometer, the intraclass correlation coefficient (ICC) varied from 0.75 to 0.96 and the overall ICC score was 0.91 (p< 0.001). In the test-retest reliability of ankle dorsiflexion ROM using a tape measure, ICC varied from 0.98 to 0.99 and the overall ICC score was 0.99 (p< 0.001). Ankle dorsiflexion ROM using a universal goniometer had a standard error of measurement (SEM) of 2.86 and a minimum detectable change (MDC) of 7.94. Ankle dorsiflexion ROM using a tape measure had an SEM of 1.01 and a MDC of 2.80.
Conclusions:
The tape measure method in a weight-bearing position was more reliable than using a universal goniometer in a non-weight-bearing position in children with CP.