The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Bougies as an aid for endotracheal intubation with the Airway Scope: bench and manikin comparison studies
Ichiro Takenaka1, Kazuyoshi Aoyama2, Tamao Iwagaki3
1Department of Anesthesia, Kyushu Rosai Hospital, 1-1 Sonekita, Kokuraminami, Kitakyushu, 800-0296, Japan. dd6xj6rx7@yahoo.co.jp.
Background:
When encountering a difficult airway with an Airway Scope (AWS) a bougie can be inserted into the endotracheal tube in the AWS channel. The angulated tip of the bougie can be guided toward the glottis by rotating it. We tested the ease of rotating bougies (Venn reusable, Boussignac, Portex single-use, and Frova) in an endotracheal tube when placed in the AWS channel.
Methods:
Bench study: Seven anesthesiologists inserted each of the four types of bougies into a 7.0 mm endotracheal tube in an AWS channel and rotated the bougie end (side of bougie operated by hand) clockwise or counterclockwise to an angle of 0°-180° in 45° increments. The rotation angle of the bougie tip (tracheal side) was measured for each bougie and the degree of force required to rotate them was examined. Manikin study: Using the same four bougies, the same seven anesthesiologists attempted to intubate a manikin that simulated a difficult airway. Success rate and time required for successful intubation were compared between the four bougies.
Results:
Bench study: The difference in the rotation angle between the bougie tip and end was significantly larger with Portex single-use and Frova bougies than with Venn reusable and Boussignac bougies (P < 0.01). The rotation angles of the tips of Venn reusable, Boussignac, Portex single-use, and Frova bougies were 145°/123° (clockwise / counterclockwise), 92°/108°, 46°/56°, and 39°/51°, respectively, when their ends were rotated to an angle of 180°. Venn reusable and Boussignac bougies could be rotated in the endotracheal tube by clinically acceptable rotational force. Manikin study: Times to intubation with Venn reusable [25 (SD, 5) s] and Boussignac bougies [35 (6) s] were significantly shorter than with Portex single-use [61 (17) s] and Frova bougies [69 (22) s] (P < 0.01). There were no significant differences in success rate between the four bougies.
Conclusions:
Venn reusable and Boussignac bougies are a useful aid for intubation with an AWS. Portex single-use and Frova bougies seem to be less suitable for this technique. Different bougies may be of varying utility when used with an AWS or airway device with an endotracheal tube channel.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...