Related Experiment Video
Updated: Feb 21, 2026

Multi-step Preparation Technique to Recover Multiple Metabolite Compound Classes for In-depth and Informative Metabolomic Analysis
Published on: July 11, 2014
HPLC-MS/MS analysis of peramivir in rat plasma: Elimination of matrix effect using the phospholipid-removal
Mingdao Lei1,2, Wei Gan2, Yongbing Sun2
1Department of Pharmacy, Jiangxi Maternal and Child Health Hospital, Nanchang, China.
Abstract:
A simple HPLC-MS/MS method has been developed for the determination of peramivir in rat plasma in the present study. The analytes were separated on a C18 column (50 × 2.1 mm, 1.7 μm) and a triple-quadrupole mass spectrometer equipped with an electrospray ionization source was applied for the detection. A phospholipid-free cartridge solid-phase extraction was used to pretreat the plasma and eliminate the endogenous phospholipid. The in-source collision-induced dissociation approach showed that this pretreatment could result in negligible ion suppression from the extracted sample and could produce cleaner samples when compared with the protein precipitation. The method was linear over the concentration range of 0.12-1200.0 ng/mL for peramivir. The method was validated and successfully applied to a pharmacokinetic study after peramivir was orally and intravenously administered to Sprague-Dawley rats.