Related Experiment Video
Updated: Feb 21, 2026

Laryngeal Mask Airway LMA Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway SGA
Published on: July 14, 2023
Airway management with Fastrach laryngeal mask versus Spritztube: a prospective randomized manikin-based study
Silvia DE Rosa1,2, Fiorenza Ferrari3, Stefano Carboni Checcacci3
1Department of Anesthesiology and Intensive Care, San Bortolo Hospital, Vicenza, Italy - derosa.silvia@ymail.com.
Background:
A new promising device, the Spritztube (ST), was developed combining the ability to perform both supraglottic ventilation and orotracheal fibreoptic intubation using the same device, allowing an easy passage from supraglottic ventilation to tracheal ventilation avoiding apnea. The present study aims to compare the speed and the subjective ease of insertion of the novel tracheal tube (Spritztube®) compared to the intubating laryngeal mask airway Fastrach™ (FT-LMA) in a simulation environment.
Methods:
Each participant received verbal instruction and practical demonstration concerning "technique of insertion" for both devices on manikin and, in a randomized order, used both devices. Time of placement (T1), time of inflation (T2), the elapsed procedural time (T3), ease of insertion, time of exchange maneuver for intubation (T4), success rates and number of attempts were recorded for each EAD.
Results:
Forty-seven participants were enrolled. The ST was judged as easier insertion than FT-LMA (P<0.001) having also a significant higher success rate than FT-LMA (P<0.001). Number of attempts for insertion was similar. Number of attempts for exchange to ETI was significantly more for FT-LMA (P<0.001). FT-LMA was applied 11 s faster than the ST (median T3 Fastrach: 13 s, ST: 24 s, P<0.001) as the exchange maneuver for ETI (median T4 FT-LMA: 13 s, ST: 24 s, P<0.001).
Conclusions:
In a manikin simulation setting, insertion and intubation with and ST was performed 11 s slower than with use of a FT-LMA.
More Related Videos
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Cardiopulmonary Resuscitation II: ACLS Airway Management
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask