Image quality of virtual monochromatic images obtained using 320-detector row CT: A phantom study evaluating the
Teppei Okamura1, Yoshitake Yamada1, Minoru Yamada2
1Department of Radiology, Keio University School of Medicine, 35 Shinanomachi, Shinjuku-ku, Tokyo, 160-8582, Japan.
Objectives:
To compare the image quality between virtual monochromatic spectral (VMS) images obtained using 320-row detector CT and polychromatic 120-kVp images reconstructed with or without iterative reconstruction using various phantom sizes.
Materials And Methods:
Torso phantoms simulating three patient sizes and containing four syringes filled with water or different contrast media (5, 10, 15mgI/mL15mgI/ml) were used. The phantoms were scanned using dual-energy (80/135-kVp) and single-energy (120-kVp) protocols at different settings (20mGy, 12mGy, and 6mGy). VMS images were generated at 1-keV intervals (range, 35-135keV). Both the VMS images and the single-energy 120-kVp images were reconstructed using filtered back projection (FBP) and adaptive iterative dose reduction 3D (AIDR-3D). The signal-to-noise ratio (SNR), and the contrast-to-noise ratio (CNR) were assessed.
Results:
Using FBP reconstruction, the SNR and CNR of the VMS images were lower than or similar to those of the 120-kVp images for most dose settings. Using AIDR-3D reconstruction, however, the 70-keV VMS images had higher SNRs and CNRs than the 120-kVp images at most settings.
Conclusions:
The image quality of VMS images with FBP reconstruction tended to be lower than that of the 120-kVp images. With the use of AIDR-3D, however, approximately 70-keV VMS images had a higher image quality than the 120-kVp images.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET