Related Experiment Video
Updated: Feb 21, 2026

The Portable Chemical Sterilizer PCS, D-FENS, and D-FEND ALL: Novel Chlorine Dioxide Decontamination Technologies for the Military
Published on: June 29, 2014
Is a chlorine dioxide wiping procedure suitable for the high-level disinfection of nasendoscopes?
Noureddine Henoun Loukili1, Nadine Lemaitre1, Benoit Guery1
1Centre Hospitalier Regional Universitaire de Lille, Lille, France.
Background:
Nasendoscopes are widely used in the outpatient ENT setting. Their reprocessing requires high-level disinfection (HLD). Recently, a wiping procedure using chlorine dioxide (ClO2) has been proposed as an alternative to HLD traditional procedures.
Objective:
To assess the effectiveness of the HLD wiping procedure versus soaking procedure on a contaminated nasendoscope.
Method:
A nasendoscope was contaminated with four strains of bacteria and Bacillus subtilis spores. After HLD either with the wiping procedure or with the soaking procedure (PA), the reduction of the initial contamination was determined.
Findings:
The wiping procedure with ClO2 displayed more than 5 log reduction for vegetative bacteria after 30 s contact time (CT) and 4 log reduction on B. subtilis spores after 2 min CT. The soaking procedure with PA displayed similar results on planktonic bacteria after 10 min CT but the log reduction of B. subtilis remained below 4.
Conclusion:
The ClO2 wiping procedure showed bactericidal and sporicidal efficacy on a contaminated nasendoscope in a shorter time compared to the PA soaking procedure. Thus, ClO2 wiping procedure might be considered as an alternative to the traditional HLD procedure for nasendoscopes.
More Related Videos
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Hand hygiene
Hand washing...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Chemical Agents for Microbial Control
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Suctioning the Nasopharyngeal Airway
Equipment Required