Clinical Implications of TiGRT Algorithm for External Audit in Radiation Oncology
Daryoush Shahbazi-Gahrouei1, Mohsen Saeb1, Shahram Monadi1
1Department of Medical Physics, School of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran.
Background:
Performing audits play an important role in quality assurance program in radiation oncology. Among different algorithms, TiGRT is one of the common application software for dose calculation. This study aimed to clinical implications of TiGRT algorithm to measure dose and compared to calculated dose delivered to the patients for a variety of cases, with and without the presence of inhomogeneities and beam modifiers.
Materials And Methods:
Nonhomogeneous phantom as quality dose verification phantom, Farmer ionization chambers, and PC-electrometer (Sun Nuclear, USA) as a reference class electrometer was employed throughout the audit in linear accelerators 6 and 18 MV energies (Siemens ONCOR Impression Plus, Germany). Seven test cases were performed using semi CIRS phantom.
Results:
In homogeneous regions and simple plans for both energies, there was a good agreement between measured and treatment planning system calculated dose. Their relative error was found to be between 0.8% and 3% which is acceptable for audit, but in nonhomogeneous organs, such as lung, a few errors were observed. In complex treatment plans, when wedge or shield in the way of energy is used, the error was in the accepted criteria. In complex beam plans, the difference between measured and calculated dose was found to be 2%-3%. All differences were obtained between 0.4% and 1%.
Conclusions:
A good consistency was observed for the same type of energy in the homogeneous and nonhomogeneous phantom for the three-dimensional conformal field with a wedge, shield, asymmetric using the TiGRT treatment planning software in studied center. The results revealed that the national status of TPS calculations and dose delivery for 3D conformal radiotherapy was globally within acceptable standards with no major causes for concern.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...