Transcatheter Aortic Valve Implantation In Patients With a Large Aortic Annulus
Peter Barr1, John Ormiston1, Jim Stewart1
1Green Lane Cardiovascular Service (Cardiology), Auckland Hospital, Auckland, New Zealand.
Background:
As the indications for transcatheter aortic valve implantation (TAVI) have expanded, so to have the demands on interventionists to allow as many patients to access this technology as possible.
Methods:
We retrospectively reviewed our TAVI database for patients who had received a 29mm SAPIEN 3 valve despite having an annular area greater than the manufacturer-recommended upper limit of 683mm2, as determined by multi-detector computed tomography (MDCT). Procedural and inpatient outcome data were collected.
Results:
The study population was 5 of 121 patients receiving a SAPIEN 3 valve since it became available in March 2015. Their annular area ranged from 691 to 800mm2. Valve deployment was successful in all patients. The deployment balloon volume was nominal, except for an additional 1ml in one patient. No patient had a new indication for permanent pacing, and no significant valvular or paravalvular regurgitation (PVR) was identified on post-procedure transthoracic echocardiography. All patients survived to hospital discharge.
Conclusions:
In this select group of patients we have demonstrated that it is safe and feasible to use the 29mm SAPIEN 3 in patients with annular dimensions greater than those recommended, with minimal balloon overfilling.