X-Bolt unforeseen placement complication: case report
A D Clarke1, Jbt Herron1, J L McVie1
1Trauma and Orthopaedics, James Cook University Hospital, South Tees Hospitals NHS Foundation Trust , Middlesbrough , UK.
Abstract:
With the introduction of the World Hip Trauma Evaluation Four clinical trial, fixation of pertrochanteric neck of femur fractures is becoming a hot topic. In this trial, the novel X-Bolt expanding bolt implant is being compared with the current gold standard of a sliding hip screw. We present a previously undescribed complication when inserting the bolt into the femoral head, where the expandable wings penetrate the femoral neck due to misplacement of the bone crusher or the X-Bolt prosthesis. This unforeseen complication required the introduction of several additional corrective intraoperative steps.