Comparing a novel hand held device (Peritron+) to standard urodynamics in measuring intravesical pressure
Vladimir A Ruzhynsky1, Stephanie Bitzos2, Ing Goping2
1University of Toronto, Toronto Western Hospital, Toronto, Ontario, Canada.
Aims:
Peritron+ is a new handheld device, which can be attached to a urethral catheter to measure intravesical pressures (Pves). The goal of this study was to assess if Pves recorded by standard urodynamics (UDs) are comparable to those recorded by the Peritron+.
Methods:
Patients with voiding dysfunction and an indication for UD underwent measurement of Pves with nonvideo UD and Peritron+. Pves were recorded at volumes of 50, 100, and 200 mL for the supine and sitting positions. A failure of the Peritron+ was considered to be a discrepancy of Pves values >3 cmH2 O between Peritron+ and UD recordings. Two-way ANOVA was used to assess for statistical significance. P-values < 0.05 were considered significant.
Results:
Ten female patients, mean age 51 years old, were enrolled in the study. The difference in Pves measurements between Peritron+ and UD were ≤3 cmH2 O at all bladder volumes measured in a supine and sitting positions. There were no adverse events and there were no malfunctions of the Peritron+ device during the study.
Conclusions:
Peritron+ handheld device accurately assessed Pves at different intravesical volumes when compared to UD measurements. The Peritron+ is a simple and safe device, which can be used in the office by a clinician and at home by a patient to monitor Pves and help to select patients who may require standard UD re-evaluation. This will be a valuable tool in managing patients with neurogenic and nonneurogenic bladders who are at a risk of upper urinary tract deterioration.
More Related Videos
Related Concept Videos
Urodynamic Studies: Uroflowmetry
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Imaging Studies VI: Voiding Cystourethrography and Cystography
Imaging Studies V: Intravenous Urography and Retrograde Pyelography
Imaging Studies II: Ultrasonography
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...