Comparative usability of modern anaesthesia ventilators: a human factors study
J Spaeth1,2, T Schweizer3, A Schmutz1,2
1Department of Anesthesiology and Critical Care, Medical Center, University of Freiburg, Hugstetter Strasse 55, 79106 Freiburg, Germany.
Background:
The anaesthesia ventilator represents the key equipment for intraoperative respiratory care. Improper operation of this device may threaten a patient's health. A self-explanatory interface facilitates handling and decreases the risk of operating errors. This study systematically evaluates the usability of user interfaces in four modern anaesthesia ventilators.
Methods:
Twenty naïve operators were asked to execute 20 tasks on each of four different anaesthesia ventilators (Avance CS2™, GE Healthcare; Flow-i™, Maquet; and Perseus™ and Primus™, Dräger) in a randomized order. The success of task execution, frequency of requests for assistance, and processing times were recorded. During the tasks, the operators' visual focus was measured via eye-tracking. Additionally, subjective assessments of usability were evaluated by a standardized questionnaire. For comparison, six experienced operators undertook the same protocol.
Results:
The overall rate of falsely executed tasks was low. Naïve operators requested assistance least when using the Perseus (26). Pooled processing times were shortest for the Perseus (222 s), followed by the Primus (223 s), the Avance (238 s), and the Flow-i (353 s). Task-specific processing times differed considerably between the devices. Eye-tracking analyses revealed associated interface issues that impeded the operators' performance. Operators rated usability best for the Perseus [mean (sd): 67 (17) arbitrary units] and worst for the Flow-i [50 (16) arbitrary units]. Results from experienced operators support these findings by trend.
Conclusions:
The usability of modern anaesthesia ventilators differs considerably. Interface issues of specific tasks impair the operator's efficiency. Eliminating the specific usability issues might improve the operator's performance and, as a consequence, the patient's safety.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...