Preventing Wedge Deformities When Treating Intertrochanteric Femur Fractures With Intramedullary Devices: A Technical Tip
Related Experiment Video
Updated: Oct 21, 2026

An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
Bennet A Butler1, Ryan S Selley1, Hobie D Summers2
1Department of Orthopedic Surgery, Northwestern Memorial Hospital, Chicago, IL.
Abstract:
Intertrochanteric femur fractures are common injuries associated with significant patient morbidity and mortality and high societal costs. Although intramedullary devices have shown promises for treating a wide range of intertrochanteric femur fractures, improper technique can lead to malalignment, fracture displacement, and/or iatrogenic fractures. In particular, a "wedge effect" has been described, in which the passage of conical reamers and the intramedullary nail itself results in the distraction of intertrochanteric fragments with lateralization of the shaft from the femoral neck and varus malalignment. Here, we describe the mechanism by which this deformity is generated and describe techniques for preventing it, including the use of a modified starting point and an alternative to the conical opening reamers.