Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
EGSnrc calculation of activity calibration factors for the Vinten ionization chamber
Abstract:
The National Physical Laboratory Vinten 671 chamber was selected as a proving ground for a new radionuclide source model in the EGSnrc software. The computational Vinten model is validated against measurements of radionuclide artifacts whose activities were determined by absolute methods. The response of the Vinten chamber is first calculated as a function of gamma energy, but more strikingly, an explicit simulation of radionuclide decay was implemented and now permits the direct determination of a calibration factor, including additional effects due to all decay paths of the radionuclide. The Monte Carlo and experimental calibration factors are found to agree at the percent level, in absolute terms.