Cardiovascular magnetic resonance imaging in patients with cardiac implantable electronic devices: a device-dependent
Sebastian Hilbert1, Cosima Jahnke1, Susanne Loebe1
1Department of Rhythmology, HELIOS Heart Center-University of Leipzig, Struempellstr 39, Leipzig, Germany.
Aims:
To prospectively determine evaluability of routine cardiovascular magnetic resonance (CMR) diagnostic modules in a referral population of implanted rhythm device all-comers, and to establish a device-dependent CMR imaging strategy to achieve optimal image quality.
Methods And Results:
One hundred and twenty-eight patients with cardiac implantable electronic devices [insertable cardiac monitoring system, n = 14; implantable loop-recorder, n = 21; pacemaker, n = 31; implantable cardioverter-defibrillator (ICD), n = 50; and cardiac resynchronization therapy defibrillator (CRT-D), n = 12] underwent clinically indicated CMR at 1.5 T. CMR protocols were tailored to the clinical indication and consisted of cine, perfusion, T1-/T2-weighted, late-gadolinium enhancement (LGE), 3D angiographic, and post-contrast cine spoiled gradient echo (SGE) scans. Image quality was determined using a 4-grade visual score per myocardial segment. Segmental evaluability was strongly influenced by device type and location with the highest proportion of non-diagnostic images encountered in the presence of ICD/CRT-D systems. Cine steady-state free-precession (SSFP) imaging was found to be mostly non-diagnostic in ICD/CRT-D patients, but a significant improvement of image quality was demonstrated when using SGE sequences with a further incremental improvement post-contrast resulting in an overall four-fold higher likelihood of achieving good image quality. LGE scans were found to be non-diagnostic in about one-third of left-ventricular segments of ICD/CRT-D patients but were artefact-free in > 94% for all other device types.
Conclusion:
Device type and location constitute the main independent predictors of CMR image quality and thus, need to be considered during protocol adaptation. Most notably, post-contrast SGE cine imaging proved superior to conventionally used SSFP sequences. Thus, following the proposed device-dependent CMR imaging strategy, diagnostic image quality can be achieved in the majority of device patients.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System V: CT
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...