Related Experiment Video
Updated: Feb 20, 2026

Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
Multicenter Comparison of Novel Self-Expanding Versus Balloon-Expandable Transcatheter Heart Valves
Oliver Husser1, Won-Keun Kim2, Costanza Pellegrini1
1Klinik für Herz und Kreislauferkrankungen, Deutsches Herzzentrum München, Technical University Munich, Munich, Germany.
Objectives:
This study sought to compare 2 next-generation transcatheter heart valves (THV), the self-expanding ACURATE neo (NEO) and the balloon-expandable SAPIEN 3 (S3), in terms of device failure and early safety at 30 days.
Background:
Deployment of these THV showed promising initial clinical results. However, no comparative data are available.
Methods:
Of 1,121 treated patients at 3 centers, a 1-to-2 nearest neighbor matching was performed to identify 2 patients treated with S3 (n = 622) for each patient treated with NEO (n = 311).
Results:
In-hospital complications were comparable between NEO and S3, including stroke (1.9% vs. 2.4%; p = 0.64), major vascular complications (10.3% vs. 8.5%; p = 0.38), or life-threatening bleeding (4.2% vs. 3.7%; p = 0.72). Device failure with NEO was comparable with S3 (10.9% vs. 9.6%; odds ratio: 1.09 [95% confidence interval: 0.69 to 1.73]; p = 0.71) with more paravalvular leakage (PVL II+, 4.8% vs. 1.8%; p = 0.01), but less elevated gradients (≥20 mm Hg, 3.2% vs. 6.9%; p = 0.02) and pacemaker implantations (9.9% vs. 15.5%; p = 0.02). Thirty-day mortality (2.3% vs. 1.9%; p = 0.74) and the early safety composite endpoint (15.8% vs. 15.6%; hazard ratio: 0.97 [95% confidence interval: 0.68 to 1.39]; p = 0.88) were similar with NEO and S3.
Conclusions:
Very high success rates were achieved for both valves, and the clinical and procedural results were comparable. Compared with S3, NEO was associated with less new pacemaker implantations and less elevated gradients, but with more paravalvular leakage.
More Related Videos
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
06:04Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025