Image noise reduction technology allows significant reduction of radiation dosage in cardiac device implantation
Rainer Hoffmann1, Lukas Langenbrink1, Doris Reimann1
1Department of Cardiology, Angiology and Sleep Medicine, Bonifatius Hospital Lingen, Lingen, Germany.
Background:
Novel x-ray systems with real-time image noise reduction technology (INRT) to reduce radiation dose during fluoroscopy and cine acquisition have become available. This study evaluated the reduction of radiation dose in device implantation with INRT.
Methods:
Radiation dose data from 132 consecutive new device implantation procedures (102 pacemaker [PM] or implantable cardioverter defibrillator [ICD] and 30 cardiac resynchronization therapy [CRT] devices) performed between January 2015 and December 2015 on an angiography system with INRT (Allura ClarityIQ) were collected. For comparison, radiation dose data from 147 consecutive device implantation procedures (121 PM/ICDs and 26 CRT devices) performed between June 2013 and September 2014 on a C-arm system with continuous and pulsed fluoroscopy option (4 frames/second) were evaluated. Total dose area product (DAP), fluoroscopy DAP, and cine DAP were evaluated.
Results:
Patient age, gender and body weight, procedure, and fluoroscopy times were similar between systems. In PM/ICD cases, DAP of INRT and C-arm system was similar (423 ± 381 cGycm2 vs 417 ± 517 cGycm) due to pulsed fluoroscopy with the C-arm system (78% of time) and sparse use of cine. In CRT procedures requiring higher image quality (82% use of continuous fluoroscopy with C-arm system), DAP of INRT was significantly lower (1,544 ± 834 cGycm vs 7,252 ± 6,431 cGycm, P < 0.001) due to less fluoroscopy DAP (1,414 ± 757 cGycm vs 5,854 ± 6,767 cGycm) and less cine DAP (130 ± 106 cGycm vs 1,399 ± 1,342 cGycm). Considering all procedures, total DAP was reduced by 60% using INRT.
Conclusion:
Novel INRT results in a substantial lowering of radiation dose in device implantation, in particular, in complex CRT implantation procedures requiring high image quality.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System IV: CMRI