Short-Term Repeatability of Noninvasive Aortic Pulse Wave Velocity Assessment: Comparison Between Methods and Devices
Andrea Grillo1,2, Gianfranco Parati1,2, Matteo Rovina3
1Department of Cardiovascular Neural and Metabolic Sciences, Istituto Auxologico Italiano, Italy.
Background:
Aortic pulse wave velocity (PWV) is an indirect index of arterial stiffness and an independent cardiovascular risk factor. Consistency of PWV assessment over time is thus an essential feature for its clinical application. However, studies providing a comparative estimate of the reproducibility of PWV across different noninvasive devices are lacking, especially in the elderly and in individuals at high cardiovascular risk.
Methods:
Aimed at filling this gap, short-term repeatability of PWV, estimated with 6 different devices (Complior Analyse, PulsePen-ETT, PulsePen-ET, SphygmoCor Px/Vx, BPLab, and Mobil-O-Graph), was evaluated in 102 high cardiovascular risk patients hospitalized for suspected coronary artery disease (72 males, 65 ± 13 years). PWV was measured in a single session twice, at 15-minute interval, and its reproducibility was assessed though coefficient of variation (CV), coefficient of repeatability, and intraclass correlation coefficient.
Results:
The CV of PWV, measured with any of these devices, was <10%. Repeatability was higher with cuff-based methods (BPLab: CV = 5.5% and Mobil-O-Graph: CV = 3.4%) than with devices measuring carotid-femoral PWV (Complior: CV = 8.2%; PulsePen-TT: CV = 8.0%; PulsePen-ETT: CV = 5.8%; and SphygmoCor: CV = 9.5%). In the latter group, PWV repeatability was lower in subjects with higher carotid-femoral PWV. The differences in PWV between repeated measurements, except for the Mobil-O-Graph, did not depend on short-term variations of mean blood pressure or heart rate.
Conclusions:
Our study shows that the short-term repeatability of PWV measures is good but not homogenous across different devices and at different PWV values. These findings, obtained in patients at high cardiovascular risk, may be relevant when evaluating the prognostic importance of PWV.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.