Pedicle Screw Placement Using an Augmented Reality Head-Mounted Display in a Porcine Model
Published on: May 24, 2024
Accuracy of a patient-specific 3D printed drill guide for placement of cervical transpedicular screws
Sinead E Hamilton-Bennett1, Bill Oxley1, Sebastien Behr1
1Willows Veterinary Centre and Referral Service, Solihull, West Midlands, United Kingdom.
Objective:
To develop a patient-specific 3-dimensional (3D) printed drill guide for placement of cervical transpedicular screws and to assess its accuracy.
Study Design:
Prospective case-series.
Sample Population:
Thirty-two cervical pedicle screws (CPS) placed in 3 large breed dogs.
Methods:
Computed tomographic (CT) data of the cervical vertebrae were exported to a medical image processing software and 3D virtual vertebral models were created for each vertebra. These models were processed in a computer aided design (CAD) software to determine the optimal trajectory and size of the CPS. Virtual drill guides were created for each patient, 3D-printed, and used intraoperatively. Locking titanium screw heads were bonded with polymethylmethacrylate cement to stabilize affected vertebral segments. Postoperative CT was used to assess the radiological accuracy of CPS placement in each dog. For each screw, CAD files were analyzed to determine a screw-diameter-to-pedicle-width-ratio (SDPWR) at the narrowest point of the pedicle.
Results:
A total of 32 CPS were placed, measuring 3.5 mm (n = 20), 2.7 mm (n = 11), and 2.4 mm (n = 1) in diameter. The majority (29/32) of these screws were placed without evidence of vertebral canal breach (grade 0), whereas a vertebral canal breach <2 mm (grade 1) was detected in 3/32 screws. This outcome was achieved despite a mean SDPWR of 0.75 (range 0.58-0.93).
Conclusion:
The use of a 3D-printed patient-specific drill guide permitted accurate placement of 32 bicortical pedicle screws in the caudal cervical vertebrae of 3 dogs. This technique may improve clinical outcome through superior biomechanical properties of screws, reduced surgical time, and reduced morbidity. These results warrant evaluation of patient outcome in a larger population.