Monoclonal protein reference change value as determined by gel-based serum protein electrophoresis
Mina Salamatmanesh1, Christopher R McCudden2, Arleigh McCurdy3
1Faculty of Health Sciences, University of Ottawa, Ottawa, Ontario, Canada; Children's Hospital of Eastern Ontario Research Institute, Ottawa, Ontario, Canada.
Introduction:
The International Myeloma Working Group recommendations for monitoring disease progression or response include quantitation of the involved monoclonal immunoglobulin. They have defined the minimum change criteria of ≧25% with an absolute change of no <5g/L for either minimal response or progression. Limited evidence is available to accurately determine the magnitude of change in a monoclonal protein to reflect a true change in clinical status. Here we determined the analytical and biological variability of monoclonal proteins in stable monoclonal gammopathy of undetermined significance (MGUS) patients.
Method:
Analytical variability (CVa) of normal protein fractions and monoclonal proteins were assessed agarose gel-based serum protein electrophoresis. Sixteen clinically stable MGUS patients were identified from our clinical hematology database. Individual biological variability (CVi) was determined and used to calculate a monoclonal protein reference change value (RCV).
Result:
Analytical variability of the normal protein fractions (albumin, alpha-1, alpha-2, beta, total gamma) ranged from 1.3% for albumin to 5.8% for the alpha-1 globulins. CVa of low (5.6g/L) and high (32.2g/L) concentration monoclonal proteins were 3.1% and 22.2%, respectively. Individual CVi of stable patients ranged from 3.5% to 24.5% with a CVi of 12.9%. The reference change value (RCV) at a 95% probability was determined to be 36.7% (low) 39.6% (high) using our CVa and CVi.
Conclusions:
Serial monitoring of monoclonal protein concentration is important for MGUS and multiple myeloma patients. Accurate criteria for interpreting a change in monoclonal protein concentration are required for appropriate decision making. We used QC results and real-world conditions to assess imprecision of serum protein fractions including low and high monoclonal protein fractions and clinically stable MGUS patients to determine CVi and RCV. The calculated RCVs of 36.7% (low) and 39.6% (high) in this study were greater that reported previously and greater than the established criteria for relapse. Response criteria may be reassessed to increase sensitivity and specificity for detection of response.