Tilt Testing with Combined Lower Body Negative Pressure: a "Gold Standard" for Measuring Orthostatic Tolerance
Published on: March 21, 2013
Validation of Central and Peripheral Non-Invasive Hemodynamic Variables Using an Oscillometric Method
Darío Echeverri1, Alejandro Pizano2, Jaime Cabrales1
1Vascular Function Research Laboratory, Interventional Cardiology Department, Fundación CardioInfantil-Instituto de Cardiología, Calle 163 A número 13B-60, Torre H. 3 Piso, Bogotá, Colombia.
The non-invasive Arteriograph system accurately measures peripheral and central blood pressure, correlating well with invasive cardiac catheterization. This validated method aids cardiovascular risk assessment and prevention strategies.
Area of Science:
- Cardiovascular Physiology
- Medical Device Validation
- Arterial Stiffness Measurement
Background:
- Peripheral and central blood pressures are key indicators of arterial stiffness and cardiovascular risk.
- Non-invasive methods for measuring these pressures are increasingly important in clinical practice.
Purpose of the Study:
- To validate the Arteriograph system, a non-invasive oscillometric device, against invasive pressure measurements obtained via cardiac catheterization.
- To assess the correlation and agreement between the Arteriograph and the gold standard invasive method for peripheral and central blood pressure.
Main Methods:
- A prospective cohort study involving 100 adult patients (mean age 64 ± 11 years).
- Simultaneous peripheral and central blood pressure measurements using the Arteriograph system and left cardiac catheterization (gold standard) via radial artery.
- Adherence to the ARTERY Society's standardized protocol for central blood pressure validation.
Main Results:
- The Arteriograph system demonstrated high correlation with invasive measurements: intraclass correlation coefficient (ICC) of 0.80 for brachial pressure and 0.91 for central pressure (p < 0.001).
- Bland-Altman analysis and linear regressions confirmed good agreement between the oscillometric and invasive methods.
- The study confirmed the reliability of the Arteriograph for both peripheral and central blood pressure estimation.
Conclusions:
- The non-invasive oscillometric Arteriograph system is a valid and easy-to-use tool for estimating hemodynamic variables, including central and peripheral arterial pressure.
- The method shows good agreement and conformity with the gold standard, applicable across diverse patient populations and conditions.
- This validated technique can enhance cardiovascular assessment, optimize treatment, and contribute to future cardiovascular prevention efforts.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
11:26Assessing Cerebral Autoregulation via Oscillatory Lower Body Negative Pressure and Projection Pursuit Regression
Published on: December 10, 2014
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.