Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Verification of junction dose between VMAT arcs of total body irradiation using a Sun Nuclear ArcCHECK phantom
Colm T Morrison1,2, Kirsty L Symons1, Simon J Woodings1,2
1Department of Radiation Oncology, Genesis Cancer Care WA, Fiona Stanley Hospital, 11 Robin Warren Drive, Murdoch, 6150, Western Australia.
Abstract:
A volumetric modulated arc therapy (VMAT) approach to total body irradiation (TBI) has recently been introduced at our institution. The planning target volume (PTV) is divided into separate sub-volumes, each being treated with 2 arcs with their own isocentre. Pre-treatment quality assurance of beams is performed on a Sun Nuclear ArcCHECK diode array. Measurement of junction regions between VMAT arcs with separate isocentres has previously been performed with point dose ionization chamber measurements, or with films. Translations of the ArcCHECK with respect to a known distance between the adjacent isocentres of two arcs, which are repeated with the ArcCHECK in an inverted position, allows the recording of a junction dose map. A 3%/3 mm global gamma analysis (10% threshold) pass rate for arc junctions were comparable to their component arcs. Dose maps of junction regions between adjacent arcs with different isocentres can be readily measured on a Sun Nuclear ArcCHECK diode array.
Related Concept Videos
Absorption of Radiation
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Nuclear Transmutation
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...