Homogeneity of wheat flour in 5ml containers for certified reference materials
R Furukawa1, Y Unno1, T Miura1
1National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology, 1-1-1 Umezono, Tsukuba, Ibaraki, Japan.
Summary
Certified reference materials were developed using wheat flour for well-type HPGe detectors. Homogeneity testing confirmed minimal sample variation, ensuring suitability for detector validation checks.
Area of Science:
- Nuclear instrumentation
- Metrology
- Analytical chemistry
Background:
- High-purity germanium (HPGe) detectors are crucial for gamma-ray spectroscopy.
- Accurate quantification requires reliable reference materials for detector calibration and validation.
- Well-type HPGe detectors necessitate specific sample geometries for optimal performance.
Purpose of the Study:
- To develop certified reference materials (CRMs) for well-type HPGe detectors.
- To establish the homogeneity of wheat flour as a potential matrix for CRMs.
- To validate the suitability of the prepared sample for detector performance checks.
Main Methods:
- Selection of wheat flour as a representative sample matrix.
- Evaluation of sample homogeneity within a 5ml well-type container.
- Calculation of homogeneity uncertainty (u_hom) using established statistical methods.
Main Results:
- Wheat flour was successfully prepared as a sample matrix.
- Homogeneity of the sample in the well-type container was assessed.
- The inhomogeneity was found to be sufficiently small (u_hom = 0.44%) for the intended application.
Conclusions:
- The developed wheat flour sample meets the homogeneity requirements for use as a reference material.
- This material is suitable for validation checks of well-type HPGe detectors.
- The study provides a foundation for the certification of these reference materials.
More Related Videos
Related Concept Videos
Test for Homogeneity
The goodness–of–fit test can be used to decide whether a population fits a given distribution, but it will not suffice to decide whether two populations follow the same unknown distribution. A different test, called the test for homogeneity, can be used to conclude whether two populations have the same distribution. To calculate the test statistic for a test for homogeneity, follow the same procedure as with the test of independence. The hypotheses for the test for homogeneity can be stated as...
Fineness of Cement
The fineness of cement directly influences the rate of hydration, as the hydration begins at the surface of the cement particles. In addition to hydration, the fineness of cement is vital for various properties of concrete including workability, gypsum requirement, and long-term behavior. The fineness of cement is represented in terms of the specific surface of cement which is typically measured in square meters per kilogram, with several methods available for this determination.