Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
The validity and inter-device variability of the Apple Watch™ for measuring maximal heart rate
Grant Abt1, James Bray1, Amanda Clare Benson2
1a School of Life Sciences , The University of Hull , Kingston upon Hull , UK.
Abstract:
Maximal heart rate (HRmax) is a fundamental measure used in exercise prescription. The Apple Watch™ measures heart rate yet the validity and inter-device variability of the device for measuring HRmax are unknown. Fifteen participants completed a maximal oxygen uptake test while wearing an Apple Watch™ on each wrist. Criterion HRmax was measured using a Polar T31™ chest strap. There were good to very good correlations between the watches and criterion (left: r = 0.87 [90%CI: 0.67 to 0.95]; right: r = 0.98 [90%CI: 0.94 to 0.99]). Standardised mean bias for the left and right watches compared to the criterion were 0.14 (90%CI: -0.12 to 0.39; trivial) and 0.04 (90%CI: -0.07 to 0.15; trivial). Standardised typical error of the estimate for the left and right watches compared to the criterion were 0.51 (90%CI: 0.38 to 0.80; moderate) and 0.22 (90%CI: 0.16 to 0.34; small). Inter-device standardised typical error was 0.46 (90%CI: 0.36 to 0.68; moderate), ICC = 0.84 (90%CI: 0.65 to 0.93). The Apple Watch™ has good to very good criterion validity for measuring HRmax, with no substantial under- or over-estimation. There were moderate and small prediction errors for the left and right watches. Inter-device variability in HRmax is moderate.
Related Concept Videos
Factors Influencing Heart Rate
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
Special considerations while measuring pulse
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...