Variations in electrode impedance during and after cochlear implantation: Round window versus extended round window
Jun Wang1, Jiaqiang Sun1, Jingwu Sun1
1Department of Otorhinolaryngology, Head and Neck Surgery, Anhui Provincial Hospital, Anhui Medical University, No.17 Lujiang road, Hefei 230001, Anhui, China.
International Journal of Pediatric Otorhinolaryngology
|November 7, 2017
Summary
Cochlear implantation using round window insertion (RWI) and extended round window insertion (ERWI) showed no significant differences in electrode impedance. Both surgical techniques are equally effective for hearing implant outcomes.
Area of Science:
- Otolaryngology
- Neurosurgery
- Biomedical Engineering
Background:
- Cochlear implantation is a common treatment for severe to profound hearing loss.
- Electrode impedance is a critical measure of cochlear implant function.
Purpose of the Study:
- To compare intra- and postoperative electrode impedances between round window insertion (RWI) and extended round window insertion (ERWI) techniques in cochlear implantation.
- To evaluate the impact of surgical approach on electrode impedance over time.
Main Methods:
- Fifty patients with congenital hearing loss underwent unilateral cochlear implantation with standard electrode arrays.
- Patients were divided into two groups: 35 with RWI and 15 with ERWI.
- Electrode impedance was measured intraoperatively, and at one week and one month postoperatively.
Main Results:
- No statistically significant differences in electrode impedance were observed between RWI and ERWI groups at any measurement point (intraoperative, 1 week, 1 month).
- Electrode impedance increased significantly from intraoperative to one-month postoperative measurements in both groups.
- No significant differences in operative duration or complications were noted between the two surgical routes.
Conclusions:
- Both RWI and ERWI are viable surgical approaches for cochlear implantation regarding electrode impedance.
- The choice between RWI and ERWI does not appear to impact postoperative electrode impedance or implant function.
- Further research may explore long-term impedance trends and functional outcomes.