Exploiting an automated microfluidic hydrodynamic sequential injection system for determination of phosphate
Wanpen Khongpet1, Somkid Pencharee2, Chanida Puangpila1
1Laboratory for Analytical Instrumentation and Electrochemistry Innovation, Department of Chemistry, Faculty of Science, Chiang Mai University, Chiang Mai 50200, Thailand.
Talanta
|November 8, 2017
Summary
A new microfluidic hydrodynamic sequential injection (μHSI) system offers a portable and cost-effective way to determine phosphate levels. This compact device uses minimal chemicals and achieves accurate results for soil sample analysis.
Area of Science:
- Analytical Chemistry
- Microfluidics
- Spectrophotometry
Background:
- Traditional methods for phosphate determination can be time-consuming and require significant reagent volumes.
- There is a need for portable, cost-effective, and low-consumptive analytical systems for field applications.
Purpose of the Study:
- To design and fabricate a microfluidic hydrodynamic sequential injection (μHSI) spectrophotometric system.
- To demonstrate the system's capability for phosphate determination using the molybdenum blue method.
- To evaluate the system's performance in terms of linearity, detection limit, and accuracy in soil samples.
Main Methods:
- Fabrication of a microfluidic channel platform using laser engraving on acrylic.
- Integration of solenoid valves for programmable liquid flow control based on HSI principles.
- Sequential aspiration of reagents (ascorbic acid, sample/standard, molybdate) and carrier water for phosphate detection via molybdenum blue spectrophotometry.
Main Results:
- A linear calibration graph was obtained for phosphate concentrations ranging from 0.1 to 6 mg P L⁻¹.
- The system achieved a detection limit of 0.1 mg L⁻¹.
- Analysis of soil samples yielded percent recoveries between 91.2% and 107.3%, showing good agreement with batch methods.
Conclusions:
- The developed μHSI system is compact, durable, portable, and cost-effective.
- The system demonstrates efficient and accurate phosphate determination with minimal chemical consumption.
- The μHSI system is suitable for on-site analysis of phosphate in environmental samples like soil.