Related Experiment Video
Updated: Feb 19, 2026

Author Spotlight: Innovative Methodology for Implanting and Securing Neural Probes in the Rodent Spinal Cord
Published on: July 12, 2024
A novel three-dimensional printed guiding device for electrode implantation of sacral neuromodulation
1Department of Gastrointestinal Surgery, Renji Hospital, School of Medicine, Shanghai Jiao Tong University, Shanghai, China.
Aim:
The aim was to test the feasibility of a novel three-dimensional (3D) printed guiding device for electrode implantation of sacral neuromodulation (SNM).
Method:
A 3D printed guiding device for electrode implantation was customized to patients' anatomy of the sacral region. Liquid photopolymer was selected as the printing material. The details of the device designation and prototype building are described. The guiding device was used in two patients who underwent SNM for intractable constipation. Details of the procedure and the outcomes are given.
Results:
With the help of the device, the test needle for stimulation was placed in the target sacral foramen successfully at the first attempt of puncture in both patients. The time to implant a tined SNM electrode was less than 20 min and no complications were observed. At the end of the screening phase, symptoms of constipation were relieved by more than 50% in both patients and permanent stimulation was established.
Conclusion:
The customized 3D printed guiding device for implantation of SNM is a promising instrument that facilitates a precise and quick implantation of the electrode into the target sacral foramen.
More Related Videos
08:26Surgical Implantation of Chronic Neural Electrodes for Recording Single Unit Activity and Electrocorticographic Signals
Published on: February 24, 2012
15:43Design and Fabrication of Ultralight Weight, Adjustable Multi-electrode Probes for Electrophysiological Recordings in Mice
Published on: September 8, 2014