Albuterol Delivery Efficiency in a Pediatric Model of Noninvasive Ventilation With Double-Limb Circuit
Jeanne Velasco1, Ariel Berlinski2
1University of Arkansas for Medical Sciences, Department of Pediatrics, Pulmonology Section, Little Rock, Arkansas.
Background:
Noninvasive ventilation (NIV) is used to treat respiratory failure in patients with concomitant need for aerosol delivery. Limited pediatric data are available on aerosol delivery efficiency, and none at all regarding aerosol delivery efficiency with a double-limb circuit. We compared the effect of position in the double-limb ventilator circuit, types of nebulizer, and ventilator settings on aerosol delivery efficiency in a pediatric model of NIV. We hypothesized that placing a vibrating mesh nebulizer at the ventilator and using the highest inspiratory pressures would increase aerosol delivery efficiency.
Methods:
A breathing simulator was connected in series to a low dead-space filter holder (lung dose) and to an anatomically correct head/airway model of a 5-year-old child. A non-vented mask connected the model to a ventilator operated on noninvasive bi-level mode and assembled with a double-limb, heated-wired adult circuit. Inspiratory pressures of either 15 or 20 cm H2O and an expiratory pressure of 5 cm H2O were used. Two different vibrating mesh nebulizers and 2 different jet nebulizers loaded with albuterol solution were studied. Albuterol was measured with spectrophotometry. Aerosol delivery efficiency was calculated as lung dose expressed as a percentage of the nominal dose.
Results:
A vibrating mesh nebulizer before the mask or Y-piece provided the highest delivery efficiency and outperformed a vibrating mesh nebulizer integrated into the mask. Vibrating mesh nebulizers delivered more drug than jet nebulizers, regardless of their position in the circuit. Increasing the inspiratory pressure only improved aerosol delivery efficiency with a jet nebulizer placed at the ventilator.
Conclusions:
In a pediatric model of NIV, the effect of nebulizer position on aerosol delivery efficiency depends on the type of device and its placement in the ventilator circuit. A vibrating mesh nebulizer placed at the mask or before the Y-piece of the double-limb circuit provided the highest aerosol drug delivery during NIV. Data generated with invasive ventilation models should not be generalized to NIV models.
More Related Videos
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
07:28Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Inhaled Medications
Cardiopulmonary Resuscitation II: ACLS Airway Management