Quantitation Error in 1H MRS Caused by B1 Inhomogeneity and Chemical Shift Displacement
Hidehiro Watanabe1, Nobuhiro Takaya1
1Center for Environmental Measurement and Analysis, National Institutes for Environmental Studies.
Purpose:
The quantitation accuracy in proton magnetic resonance spectroscopy (1H MRS) improves at higher B0 field. However, a larger chemical shift displacement (CSD) and stronger B1 inhomogeneity exist. In this work, we evaluate the quantitation accuracy for the spectra of metabolite mixtures in phantom experiments at 4.7T. We demonstrate a position-dependent error in quantitation and propose a correction method by measuring water signals.
Materials And Methods:
All experiments were conducted on a whole-body 4.7T MR system with a quadrature volume coil for transmission and reception. We arranged three bottles filled with metabolite solutions of N-acetyl aspartate (NAA) and creatine (Cr) in a vertical row inside a cylindrical phantom filled with water. Peak areas of three singlets of NAA and Cr were measured on three 1H spectra at three volume of interests (VOIs) inside three bottles. We also measured a series of water spectra with a shifted carrier frequency and measured a reception sensitivity map.
Results:
The ratios of NAA and Cr at 3.92 ppm to Cr at 3.01 ppm differed amongst the three VOIs in peak area, which leads to a position-dependent error. The nature of slope depicting the relationship between peak areas and the shifted values of frequency was like that between the reception sensitivities and displacement at every VOI.
Conclusion:
CSD and inhomogeneity of reception sensitivity cause amplitude modulation along the direction of chemical shift on the spectra, resulting in a quantitation error. This error may be more significant at higher B0 field where CSD and B1 inhomogeneity are more severe. This error may also occur in reception using a surface coil having inhomogeneous B1. Since this type of error is around a few percent, the data should be analyzed with greater attention while discussing small differences in the studies of 1H MRS.
Related Concept Videos
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...
NMR Spectrometers: Resolution and Error Correction
NMR Spectroscopy: Chemical Shift Overview
For instance, the proton...
¹H NMR Chemical Shift Equivalence: Homotopic and Heterotopic Protons
¹H NMR of Conformationally Flexible Molecules: Temporal Resolution
¹H NMR: Complex Splitting
Splitting diagrams or splitting tree diagrams are routinely used to depict such complex couplings. While drawing splitting diagrams, the splitting with the larger coupling constant is usually applied...