Related Experiment Video
Updated: Feb 19, 2026

Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Failure of an expandable cage-plate cervical vertebral body replacement: case report of a device related complication
1Department of Orthopaedics Klinikum Wels-Grieskirchen, Grieskirchnerstr. 42, 4600 Wels, Austria. Klemens.trieb@klinikum-wegr.at.
Abstract:
Instabilities of the cervical spine after complete or incomplete corpectomy may be the result from resection due to stenosis induced myelopathy, tumors, infections or fractures. In this article, we report a 49-year-old female patient after surgical decompression and stabilization at C5/7 with permanent cervicobrachialgia due to cervical stenosis at C5/7. Corpectomy of C6 and stabilization of C5/7 were performed with a combination of cage and plate. At an outpatient control 16 months after reoperation, the patient complained of neck pain, while the control X-ray showed a fracture of the implant. Due to instability of the device, a revision was indicated, resulting in removal of all components with restabilization. Definitive stabilization was performed with a PINA® cage, discectomy and cage implantation at C4/5 and C7/Th1 and a plate from C4 to Th1 with screw fixation in C4/5/Th1. Postoperative course was complication free and the patient was without pain or neurological symptoms at follow-up after 12 months with all implants in radiologically regular position. To the best of our knowledge, this is the first case report to describe such a complication.

