Validation of an Ergonomic Method of Withdrawing 99mTc-Radiopharmaceuticals
Sandy Blondeel-Gomes1, Solène Marie2, Julien Fouque3
1Department of Radio-Pharmacology, Institut Curie Hospital, Paris, France; and sandy.blondeel@curie.fr.
Abstract:
The main objective of the present work was to ensure the quality of radiopharmaceutical syringes withdrawn with a system consisting of a spinal needle and an obturator IN-Stopper. Methods: Visual examinations and physicochemical tests were performed at baseline and 4 h for 99mTc-albumin nanocolloid and at baseline and 7 h for 99mTc-eluate, 99mTc-hydroxymethylene diphosphonate, and 99mTc-human serum albumin. Microbiologic validation was performed according to the European pharmacopoeia. Fingertip radiation exposure was evaluated to confirm the safety of the system. Results: The results showed stable visual and physicochemical properties. The integrity of the connector was not affected after 30 punctures (no cores). No microbiologic contamination was found on tested syringes. Concerning radiation safety, no overexposure was reported with the system. Conclusion: The system could be used 30 times. The stability of syringes withdrawing radiopharmaceuticals with this method is guaranteed up to 4 h for 99mTc-albumin nanocolloid and 7 h for 99mTc-eluate, 99mTc-hydroxymethylene diphosphonate, and 99mTc-human serum albumin.
More Related Videos
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...