Non-enzymatic electrochemical detection of glucose with a disposable paper-based sensor using a cobalt
Sudkate Chaiyo1, Eda Mehmeti2, Weena Siangproh3
1Electrochemistry and Optical Spectroscopy Center of Excellence (EOSCE), Department of Chemistry, Chulalongkorn University, 254 Phayathai Road, Patumwan, Bangkok, Thailand.
Abstract:
We introduce for the first time a paper-based analytical device (PAD) for the non-enzymatic detection of glucose by modifying a screen-printed carbon electrode with cobalt phthalocyanine, graphene and an ionic liquid (CoPc/G/IL/SPCE). The modifying composite was characterized by UV-visible spectroscopy, energy dispersive X-ray spectroscopy (EDX), transmission electron microscopy (TEM) and scanning electron microscopy (SEM). The disposable devices show excellent conductivity and fast electron transfer kinetics. The results demonstrated that the modified electrode on PADs had excellent electrocatalytic activity towards the oxidation of glucose with NaOH as supporting electrolyte (0.1M). The oxidation potential of glucose was negatively shifted to 0.64V vs. the screen-printed carbon pseudo-reference electrode. The paper-based sensor comprised a wide linear concentration range for glucose, from 0.01 to 1.3mM and 1.3-5.0mM for low and high concentration of glucose assay, respectively, with a detection limit of 0.67µM (S/N = 3). Additionally, the PADs were applied to quantify glucose in honey, white wine and human serum. The disposable, efficient, sensitive and low-cost non-enzymatic PAD has great potential for the development of point-of-care testing (POCT) devices that can be applied in healthcare monitoring.
More Related Videos
07:51Development and Functionalization of Electrolyte-Gated Graphene Field-Effect Transistor for Biomarker Detection
Published on: February 1, 2022
11:18Manufacturing of a Nafion-coated, Reduced Graphene Oxide/Polyaniline Chemiresistive Sensor to Monitor pH in Real-time During Microbial Fermentation
Published on: January 7, 2019