Related Experiment Video
Updated: Feb 19, 2026

Diaphragmatic Ultrasound in Adults: Image Acquisition and Interpretation
Published on: January 31, 2025
Portable abdomen radiography: moving to thickness-based protocols
Adrian A Sánchez1, Ingrid Reiser2, Tina Baxter2
1Department of Radiology, The University of Chicago Medical Center, 5841 S. Maryland Ave., MC2026, Chicago, IL, 60637, USA. adrianalberto@uchicago.edu.
Background:
Default pediatric protocols on many digital radiography systems are configured based on patient age. However, age does not adequately characterize patient size, which is the principal determinant of proper imaging technique. Use of default pediatric protocols by inexperienced technologists can result in patient overexposure, inadequate image quality, or repeated examinations.
Objective:
To ensure diagnostic image quality at a well-managed patient radiation exposure by transitioning to thickness-based protocols for pediatric portable abdomen radiography.
Materials And Methods:
We aggregated patient thickness data, milliamperes (mAs), kilovoltage peak (kVp), exposure index (EI), source-to-detector distance, and grid use for all portable abdomen radiographs performed in our pediatric hospital in a database with a combination of automated and manual data collection techniques. We then analyzed the database and used it as the basis to construct thickness-based protocols with consistent image quality across varying patient thicknesses, as determined by the EI.
Results:
Retrospective analysis of pediatric portable exams performed at our adult-focused hospitals demonstrated substantial variability in EI relative to our pediatric hospital. Data collection at our pediatric hospital over 4 months accumulated roughly 800 portable abdomen exams, which we used to develop a thickness-based technique chart.
Conclusion:
Through automated retrieval of data in our systems' digital radiography exposure logs and recording of patient abdomen thickness, we successfully developed thickness-based techniques for portable abdomen radiography.
More Related Videos
Related Concept Videos
Assessment of the Abdomen III: Palpation
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...
X-ray Imaging
Assessment of the Abdomen I: Inspection and Auscultation
The abdominal examination is a cornerstone of clinical medicine, serving as a critical tool in diagnosing various gastrointestinal (GI) diseases. It involves a systematic approach that includes inspection and auscultation, each with distinct yet complementary roles in assessing the abdomen. This article will delve into these two primary methods healthcare professionals use to examine the abdomen.
Inspection of the Abdomen
The first step in any abdominal examination is inspection....
Radiological Investigation I: X-ray and CT