Aerrow: A probe-format graphite calorimeter for absolute dosimetry of high-energy photon beams in the clinical environment
Related Experiment Video
Updated: Feb 18, 2026

Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
James Renaud1, Arman Sarfehnia1,2, Julien Bancheri1
1Medical Physics Unit, McGill University, Montréal, QC, Canada.
Purpose:
In this work, the design, operation, initial experimental evaluation, and characterization of a small-scale graphite calorimeter probe - herein referred to as the Aerrow - developed for routine use in the clinical environment, are described. Similar in size and shape to a Farmer type cylindrical ionization chamber, the Aerrow represents the first translation of calorimetry intended for direct use by clinical physicists in the radiotherapy clinic.
Methods:
Based on a numerically optimized design obtained in previous work, a functioning Aerrow prototype capable of two independent modes of operation (quasi-adiabatic and isothermal) was constructed in-house. Reference dose measurements were performed using both Aerrow operation modes in a 6 MV photon beam and were directly compared to results obtained with a calibrated reference-class ionization chamber. The Aerrow was then used to quantify the absolute output of five clinical linac-based photon beams (6 MV, 6 MV FFF, 10 MV, 10 MV FFF, and 15 MV; 63.2% < %dd(10)× < 76.3%). Linearity, dose rate, and orientation dependences were also investigated.
Results:
Compared to an ion chamber-derived dose to water of 76.3 ± 0.7 cGy, the average doses measured using the Aerrow were 75.6 ± 0.7 and 74.7 ± 0.7 cGy/MU for the quasi-adiabatic and isothermal modes, respectively. All photon beam output measurements using the Aerrow in water-equivalent phantom agreed with chamber-based clinical reference dosimetry data within combined standard uncertainties. The linearity of the Aerrow's response was characterized by an adjusted R2 value of 0.9998 in the dose range of 80 cGy to 470 cGy. For the dose-rate dependence, no statistically significant effects were observed in the range of 0.5 Gy/min to 5.4 Gy/min. A relative photon beam quality dependence of 1.7% was calculated in the range of 60 Co to 24 MV (58.4% < %dd(10)× < 86.8%) using Monte Carlo. Finally, the angular dependence (gantry stationary and detector rotated) of the Aerrow's response was found to be insignificant to within ±0.5%.
Conclusions:
This work demonstrates the feasibility of using an ion chamber-sized calorimeter as a practical means of measuring absolute dose to water in the radiotherapy clinic. The potential introduction of calorimetry as a mainstream device into the clinical setting is powerful, as this fundamental technique has formed the basis of absorbed dose standards in many countries for decades and could one day form the basis of a new local absorbed dose standard for clinics.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...