Experience with the artificial urinary sphincter model AS800 in 148 patients
I J Fishman1, R Shabsigh, F B Scott
1Scott Department of Urology, Baylor College of Medicine, St. Luke's Episcopal Hospital, Houston, Texas.
The Journal of Urology
|February 1, 1989
Summary
The AS800 artificial urinary sphincter effectively restored urinary control in 90% of patients. Device removal was infrequent, primarily due to infection or erosion, highlighting its safety and efficacy in managing urinary incontinence.
Area of Science:
- Urology
- Medical Devices
- Reconstructive Surgery
Background:
- Urinary incontinence affects a significant patient population, necessitating effective treatment options.
- Artificial urinary sphincters are a key therapeutic modality for managing severe incontinence.
- The AS800 represents an advancement in artificial urinary sphincter technology.
Purpose of the Study:
- To evaluate the efficacy and safety of the AS800 artificial urinary sphincter in patients with diverse etiologies of urinary incontinence.
- To assess the device's performance, including success rates and reasons for explantation.
- To analyze the impact of the deactivation feature on patient management.
Main Methods:
- A prospective study involving 148 patients (112 male, 36 female) implanted with the AS800 artificial urinary sphincter.
- Cuff placement evaluated at the bladder neck (78 patients) or bulbar urethra (70 patients).
- Follow-up ranged from 3 to 37 months (average 20.8 months) to assess outcomes and complications.
Main Results:
- Socially acceptable urinary control was achieved in 90% of patients with active devices.
- Sphincter removal was required in 7.4% of patients, predominantly due to infection and erosion (5.4%).
- Higher success rates were observed in patients with post-surgical incontinence (failed anti-stress operations, TURP, or prostatectomy), while higher failure rates occurred in cases secondary to pelvic fracture or exstrophy/epispadias.
Conclusions:
- The AS800 artificial urinary sphincter demonstrates high efficacy and acceptable safety profiles for treating various forms of urinary incontinence.
- The device's deactivation feature offers significant benefits for catheterization, cystoscopy, and nocturnal decompression.
- Patient selection based on etiology is crucial for optimizing outcomes with the AS800 artificial urinary sphincter.