Underrecognized entity of the transient rise in the atrial capture threshold early after dual-chamber pacemaker
Yoshiko Uehara1, Kentaro Yoshida1,2, Akira Kimata2
1Department of Cardiology, Ibaraki Prefectural Central Hospital, Kasama, Japan.
Background:
Steroid-eluting pacemaker leads suppress acute rises in pacing threshold by preventing inflammatory processes. However, we occasionally encounter not persistent but transient rise in the atrial capture threshold (TRACT) early after pacemaker implantation. We believe that this phenomenon is underrecognized in clinical practice and may potentially lead to unnecessary reintervention. We aimed to clarify the prevalence, predictors, and possible mechanisms of TRACT.
Methods And Results:
We reviewed clinical records from 239 consecutive patients who underwent dual-chamber pacemaker implantation for sick sinus syndrome (SSS) (N = 102) or atrioventricular block (AVB) (N = 137). Atrial capture threshold was measured at implantation and 7 days, 2 months, and 8 months postimplantation. TRACT was defined as a rise in the threshold at day 7 to ≥twice that at implantation, with an absolute value ≥1.0 V/0.4 ms, and full recovery by 8 months into follow-up. TRACT was observed in 15 patients (6%), of whom13 (87%) suffered from SSS but not AVB. Patients with TRACT had greater body mass index (BMI) (25 ± 5 kg/m2 vs 23 ± 4 kg/m2 , P = 0.01), larger left atrium (42 ± 5 mm vs 38 ± 7 mm, P = 0.03), and were more likely to suffer from paroxysmal atrial fibrillation (60% vs 31%, P = 0.02) than those without TRACT. In multivariable logistic regression analysis, BMI and SSS were the independent predictors of TRACT (odds ratio [OR], 1.172; 95% confidence interval [CI], 1.019-1.349; P = 0.03 and OR, 11.53; 95% CI, 2.010-66.21; P = 0.006, respectively).
Conclusions:
The distinct phenomenon of TRACT was not rare in clinical practice early after dual-chamber pacemaker implantation, and its occurrence was strongly associated with SSS.