Biocompatibility and biodegradation studies of a commercial zinc alloy for temporary mini-implant applications
M Bobby Kannan1, Corey Moore2, Shruti Saptarshi3
1Biomaterials and Engineering Materials (BEM) Laboratory, James Cook University, Townsville, Queensland, 4811, Australia. bobby.mathan@jcu.edu.au.
Abstract:
In this study, the biocompatibility and in vitro degradation behaviour of a commercial zinc-based alloy (Zn-5 Al-4 Mg) were evaluated and compared with that of pure zinc for temporary orthopaedic implant applications. Biocompatibility tests were conducted using human alveolar lung epithelial cells (A549), which showed that the zinc alloy exhibits similar biocompatibility as compared to pure zinc. In vitro degradation evaluation was performed using weight loss and electrochemical methods in simulated body fluid (SBF) at 37 °C. Weight loss measurements revealed that the degradation of the zinc alloy was slightly lower during the initial immersion period (1-3 days), but marginally increased after 5 and 7 days immersion as compared to pure zinc. Potentiodynamic polarisation experiments showed that the zinc alloy exhibits higher degradation rate than pure zinc. However, electrochemical impedance spectroscopy analysis suggests that pure zinc is susceptible to localized degradation, whereas the zinc alloy exhibited passivation behaviour. Post-degradation analysis revealed localized degradation in both pure zinc and the zinc alloy.