Related Experiment Video
Updated: Feb 18, 2026

Retinal Pigment Epithelium Transplantation in a Non-human Primate Model for Degenerative Retinal Diseases
Published on: June 14, 2021
Complications Assoicated with MIRAgel for Treatment of Retinal Detachment
Miin Roh1, Nahyoung Grace Lee1, John B Miller1
1a Department of Ophthalmology , Massachusetts Eye and Ear, Harvard Medical School , Boston MA , USA.
Abstract:
MIRAgel (MIRA, Waltham, MA) is a hydrogel buckle that was introduced in 1979 as a new scleral implant for the treatment of retinal detachment. Long-term follow-up of more than 10 years revealed that the hydrolysis of the synthetic hydrophilic material leads to marked expansion of the substance, causing complications such as buckle extrusion and intrusion, eye motility disorder, cosmetic deformities, and periocular infections. Removal of the implant is the treatment of choice in cases with complications, but it is technically difficult due to the friability of the implant, severe scleral ectasia, and relatively high rate of redetachment after removal.
Insights
The MIRAgel hydrogel buckle, used for retinal detachment, can expand over time causing serious complications. Implant removal is difficult and has a high rate of re-detachment.
Area of Science:
- Ophthalmology
- Biomaterials Science
Background:
- The MIRAgel hydrogel buckle was developed in 1979 as a scleral implant for retinal detachment treatment.
- Long-term studies indicate potential complications associated with its use.
Purpose of the Study:
- To evaluate the long-term outcomes and complications of the MIRAgel hydrogel buckle in retinal detachment surgery.
Main Methods:
- Analysis of long-term follow-up data (over 10 years) of patients treated with MIRAgel implants.
- Review of complication types and outcomes following implant removal.
Main Results:
- Hydrolysis of the synthetic hydrophilic material causes significant expansion.
- Complications include buckle extrusion/intrusion, motility disorders, cosmetic deformities, and infections.
- Implant removal is challenging due to material friability, scleral ectasia, and high redetachment rates.
Conclusions:
- The MIRAgel hydrogel buckle is associated with significant long-term complications due to material degradation.
- Surgical removal of complicated implants presents technical difficulties and risks of recurrent retinal detachment.