Experimental conditions for surface microanalysis with reflection electron energy-loss spectroscopy
1Department of Materials Science & Engineering, SUNY, Stony Brook 11794-2275.
Abstract:
Experimental conditions for obtaining the optimum signal-to-background (S/B) ratio in reflection electron energy-loss spectroscopy (REELS) are investigated. It is shown that the S/B ratio can be improved by lowering the incident energy of the electrons. The spectra taken from the GaAs (660) specular reflection spot under the surface resonance condition is demonstrated to have the best S/B ratio and lowest surface plasma excitation, which is capable of providing structural information on the top few atomic layers.
More Related Videos
09:00Experimental Methods for Spin- and Angle-Resolved Photoemission Spectroscopy Combined with Polarization-Variable Laser
Published on: June 28, 2018
07:24Quantitative Atomic-Site Analysis of Functional Dopants/Point Defects in Crystalline Materials by Electron-Channeling-Enhanced Microanalysis
Published on: May 10, 2021
Related Concept Videos
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Preparation of Samples for Electron Microscopy
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Instrumentation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Lab