Non-invasive blood pressure measurement in ferrets (Mustela putorius furo) using high definition oscillometry
Y R A van Zeeland1, A Wilde1, I H Bosman1
1Department of Clinical Sciences of Companion Animals, Faculty of Veterinary Medicine, Utrecht University, Yalelaan 108, 3584 CM Utrecht, The Netherlands.
Abstract:
This study was conducted to validate the use of high definition oscillometry (HDO) for non-invasive blood pressure (NIBP) measurements in ferrets and to establish reference ranges for NIBP in minimally sedated, healthy, young adult ferrets (<4years of age). The bias, limits of agreement and correlation for HDO compared to direct arterial blood pressure (DABP) measurement were established in 14 anaesthetised ferrets. Measurements were performed at the forelimb, hind limb and tail under hypo- (<110mmHg), normo- (110-170mmHg) and hypertensive (>170mmHg) conditions. Although HDO correlated well with DABP (r>0.90), it showed significant proportional bias, whereby HDO generally underestimated DABP with hyper- and normotensive conditions, and overestimated DABP with hypotensive conditions. Measurements obtained from the hind limb showed higher bias than those obtained from the tail or forelimb (P<0.001). Based on the above, and for practical reasons, the tail was selected as the preferred site of cuff placement in ferrets. Subsequently, a cross-over study was performed in 10 ferrets to establish the minimum dose of butorphanol and midazolam needed to successfully obtain NIBP in 100% of cases. Using this dose (0.2mg/kg IM, each), reference intervals for NIBP from 63 healthy, young adult ferrets were established at 95-155mmHg (systolic), 69-109mmHg (mean) and 51-87mmHg (diastolic) arterial pressures.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...