Deviations in frequency and mode of vibration in whole-body vibration training devices with long-term and regular use
Tobias S Kaeding1, Sanam Moghaddamnia2, Momme Kück1
1Institute of Sports Medicine, Hannover Medical School (MHH), Carl-Neuberg-Straße 1, 30625 Hannover, Germany.
Abstract:
Research regarding whole body vibration training (WBVT) and its practical use may be hindered by the fact that WBVT devices generate frequencies and/or modes of vibration different from their preset adjustments. This research aimed to clarify whether prolonged regular use can generate such deviations in frequency and mode of vibration. Three WBVT devices, each used for approximately 13 months in two research projects, were tested with an accelerometer before start of the 1st study, after four months, and after 13 months (the completion of the 2nd study). Divergences between the preset and measured frequencies were calculated for all measurements. Furthermore, the total harmonic distortion (THD), an index for signal deviations from a perfect sine wave, and the sum signal-to-modulation-noise-ratio (SMNR), an indicator of fidelity, were recorded. One device had a significantly larger machine run time than the other two, and it displayed the most pronounced signs of impaired function concerning frequency, mode of vibration, and random variability (SMNR) after prolonged use. These results indicate that prolonged use will result in divergences between the preset and actual applied frequencies as well as in the mode of vibration and other accuracy measurements.
Related Concept Videos
Muscle Stimulation Frequency
Wave summation
At low firing rates, motor neurons induce individual twitch contractions in muscle fibers. These twitches...
Modes of Standing Waves - I
Fatigue
Discrete Fourier Transform
Vibrating Concrete
Forced Oscillations