Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Mixing of Head-Stem Components in Total Hip Arthroplasty
Ulrike Mueller1, Benjamin Panzram1, Steffen Braun1
1Laboratory of Biomechanics and Implant Research, Clinic for Orthopedics and Trauma Surgery, Heidelberg University Hospital, Heidelberg, Germany.
Background:
Implant manufacturers proclaim that the tapers of modular total hip arthroplasty are not standardized and can vary from manufacturer to manufacturer. That is why the combination of various components from different manufacturers ("Mix and Match") is not permitted. In this study, different taper combinations were investigated experimentally to assess the effect of "Mix and Match" regarding the taper connection strength.
Methods:
Torque-off tests using hip stems and metal femoral heads from 6 different implant manufacturers were performed. First the components were tested as intended and afterwards the stems were combined with metal heads from other manufacturers.
Results:
There was no significant difference in taper connection strength when stems from the manufacturers Link, Smith & Nephew, and Zimmer were combined with heads from other manufacturers. The Biomet stems showed a significantly reduced taper connection strength if femoral heads of Aesculap, DePuy, or Smith & Nephew were used. On the contrary, the DePuy stems in combination with the originally intended femoral heads showed a significantly lower taper connection strength compared to the use of heads from Link, Biomet, and Zimmer. The same was observed for the Aesculap stems in combination with Zimmer heads.
Conclusion:
The results of this study suggest that mixing components from different manufacturers may affect the taper connection strength and could reduce the stability. As safety should be a high priority in patient treatment, any potential risks should be avoided. Therefore, mixing and matching of heads and femoral stems from different manufacturers cannot be recommended.