Operating Room Traffic Increases Aerosolized Particles and Compromises the Air Quality: A Simulated Study
Maryam Rezapoor1, Abtin Alvand1, Elzbieta Jacek1
1The Rothman Institute at Thomas Jefferson University, Philadelphia, Pennsylvania.
The Journal of Arthroplasty
|November 28, 2017
Summary
Operating room (OR) traffic, including personnel and door openings, significantly increases particle density. Laminar airflow (LAF) effectively mitigates this particle increase, crucial for preventing surgical site infections.
Area of Science:
- Surgical Infection Prevention
- Operating Room Air Quality Management
- Biomedical Engineering
Background:
- Periprosthetic joint infection (PJI) prevention relies on minimizing bacterial contamination in the operating room (OR).
- OR traffic, specifically personnel movement and door openings, is a known factor influencing air quality.
- Quantitative data on the impact of personnel and door openings on OR particle density is limited.
Purpose of the Study:
- To quantitatively assess the influence of personnel and door openings on OR particle density.
- To evaluate the effectiveness of laminar airflow (LAF) in reducing particle density under varying traffic conditions.
Main Methods:
- Simulated study conducted in an arthroplasty OR with and without LAF.
- Particle counts measured with increasing numbers of personnel entering the room.
- Particle counts measured with increasing door opening frequency per minute.
Main Results:
- Both personnel presence and door openings significantly correlated with increased OR particle density.
- Laminar airflow (LAF) substantially reduced particle density associated with personnel and door openings.
- With LAF, particle density per person decreased from 211.19 to 18.19 particles/ft³ (P < .001).
- With LAF, particle density per door opening rate decreased from 117.80 to 1.90 particles/ft³ (P = .017).
Conclusions:
- Personnel and door openings are major contributors to OR particle load.
- Controlling OR traffic is critical for reducing particle counts and preventing PJI.
- Laminar airflow (LAF) provides a protective effect against increased particle density from OR traffic.
Related Concept Videos
Transmission-based Precautions II: Airborne and Protective Environment
2.0K
Transmission-based precautions are for patients infected or suspected to be infected (or colonized) with organisms posing a significant risk to others. The transmission precautions include airborne and protective environment precautions.
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
2.0K
Factors Affecting Pulmonary Ventilation
3.3K
Besides the pressure difference between the external environment and the lungs, the airflow rate and ease of pulmonary ventilation are also influenced by three other factors: surface tension of the fluid in the alveoli, compliance of the lungs, and airway resistance.
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
3.3K