Does the endoscope control head need to be disinfected between examinations?
G M Sobala1, C Lincoln, A T Axon
1Gastroenterology Unit, General Infirmary, Leeds.
Endoscope control heads can become contaminated during procedures like gastroduodenoscopy and colonoscopy. A simple disinfection method using detergent, alcohol, and chlorhexidine effectively cleans these potential sources of cross-infection.
Area of Science:
- Gastroenterology
- Infection Control
- Medical Device Disinfection
Background:
- Endoscopic procedures like gastroduodenoscopy and colonoscopy are essential diagnostic tools.
- Contamination of medical equipment, particularly endoscopes, poses a significant risk for healthcare-associated infections.
- The control head of endoscopes is a frequently handled component, raising concerns about its potential role in pathogen transmission.
Purpose of the Study:
- To determine the incidence of contamination on endoscope control heads after gastroduodenoscopy and colonoscopy.
- To evaluate the effectiveness of a straightforward disinfection protocol for endoscope control heads.
Main Methods:
- Assessed contamination rates of endoscope control heads during 17 gastroduodenoscopy and colonoscopy procedures.
- Artificially contaminated control heads were subjected to a disinfection process involving scrubbing with detergent and wiping with 70% isopropyl alcohol and 2% chlorhexidine swabs.
Main Results:
- Contamination of endoscope control heads was observed in 4 out of the 17 examinations conducted.
- The disinfection technique proved effective in adequately disinfecting heavily contaminated control heads.
Conclusions:
- Endoscope control heads can become contaminated during routine endoscopic examinations.
- The evaluated disinfection method is effective in eliminating contamination from endoscope control heads.
- Endoscope control heads should be considered a potential source of cross-infection and must be included in all cleaning and disinfection protocols.
More Related Videos
Related Concept Videos
Hand hygiene
Hand washing...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Transmission-based Precautions I: Contact, Enteric, and Droplets
Contact Precautions:
Contact precautions are the measures taken to prevent the transmission of infectious agents, especially epidemiologically important microorganisms such as MRSA or influenza, primarily transmitted through direct or indirect contact with an...
Methods for Controlling Microbial Growth