Abstract:
The authors devised a new technique for measuring cerebrospinal fluid movement in the spine. With gated list mode acquisition of radioactive emissions, bulk flow, oscillations, and dispersion of a bolus of isotope can be measured. The accuracy of the technique was measured in a simulation in which fluid oscillated at 71 cycles/minute and displaced at 6.5 cm/minute. With this method, the measured oscillation frequency and amplitude and the rate of translation were within 10% of actual values.