Validation of whole room indirect calorimeters: refinement of current methodologies
Russell Rising1, Thomas Foerster2, Avigdor D Arad3
1New York Obesity/Nutrition Research Center, Department of Medicine, Columbia University, New York, New York rr2986@cumc.columbia.edu.
Abstract:
Whole room indirect calorimeter (WRIC) validation techniques consist of propane combustion (PC) or infusion of mixed carbon dioxide (CO2) and nitrogen (N2) by a precision blender (PB). To determine the best method, PC of 6, 10, 22-h and PB infusions of 6, 10, and 14-h, were conducted. The 14-h infusion consisted of two metabolic settings. Energy expenditure (EE; kJ), ventilation (V; liters/min) of oxygen (VO2), VCO2, and respiratory quotient (VCO2/VO2) obtained from the WRIC were extrapolated to the respective test durations and compared to similarly calculated values. Moreover, accurate equations (AE) were derived to correct infusions for additional N2 As a final evaluation of a PC validated WRIC, weight maintenance (WM), energy balance (EB), respiratory quotient (RQ), and food quotients (FQ) were determined in 22 subjects who had repeat 24-h EE measurements. Statistical analyses (P < 0.05) were conducted (SPSS, version 23). Significant differences in RQ existed between PC and stoichiometry after 6-h. Errors for the rest of the PC tests ranged from -1.5 ± 2.4 (VCO2) to 2.8 ± 4.6% (EE). When compared with the WRIC, all uncorrected metabolic parameters for six and 10-h PB infusions were significantly different with errors from -12.8 ± 1.6 (VO2) to 6.0 ± 2.8% (RQ). The AE reduced the magnitude of errors to -12.4 ± 1.5 (RQ) to 2.2 ± 3.0% (RQ). The PB infusion with two settings showed similar performance. No differences in WM, EB, RQ, or FQ existed in the subjects. In conclusion, 10-h PC tests are sufficient for validating WRICs.
More Related Videos
04:54Using a Combination of Indirect Calorimetry, Infrared Thermography, and Blood Glucose Levels to Measure Brown Adipose Tissue Thermogenesis in Humans
Published on: June 2, 2023
10:22Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
Related Concept Videos
Constant Volume Calorimetry
Calorimetry
Constant Pressure Calorimetry
Quantifying Heat
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...