Related Experiment Video
Updated: Feb 17, 2026

Using Retinal Imaging to Study Dementia
Published on: November 6, 2017
Reproducibility of Retinal Microvascular Traits Decoded by the Singapore I Vessel Assessment Software Across the
Qi-Fang Huang1,2, Fang-Fei Wei1, Zhen-Yu Zhang1
1Studies Coordinating Centre, Research Unit Hypertension and Cardiovascular Epidemiology, KU Leuven Department of Cardiovascular Sciences, University of Leuven, Belgium.
Background:
Retinal microvascular traits predict adverse health outcomes. The Singapore I Vessel Assessment (SIVA) software improved automated postprocessing of retinal photographs. In addition to microvessel caliber, it generates measures of arteriolar and venular geometry. Few studies addressed the reproducibility of SIVA measurements across a wide age range.
Methods:
In the current study, 2 blinded graders read images obtained by nonmydriatic retinal photography twice in 20 11-year-old children, born prematurely (n = 10) or at term (n = 10) and in 60 adults (age range, 18.9-86.1 years).
Results:
Former preterm compared with term children had lower microvessel diameter and disorganized vessel geometry with no differences in intraobserver and interobserver variability. Among adults, microvessel caliber decreased with age and blood pressure and arteriolar geometry was inversely correlated with female sex and age. Intraobserver differences estimated by the Bland-Altman method did not reach significance for any measurement. Across measurements, median reproducibility (RM) expressed as percent of the average trait value was 8.8% in children (median intraclass correlation coefficient [ICC], 0.94) and 8.0% (0.97) in adults. Likewise, interobserver differences did not reach significance with RM (ICC) of 10.6% (0.85) in children and 10.4% (0.93) in adults. Reproducibility was best for microvessel caliber (intraobserver/interobserver RM, 4.7%/6.0%; ICC, 0.98/0.96), worst for venular geometry (17.0%/18.8%; 0.93/0.84), and intermediate for arteriolar geometry (10.9%/14.9%; 0.95/0.86).
Conclusions:
SIVA produces repeatable measures of the retinal microvasculature in former preterm and term children and in adults, thereby proving its usability from childhood to old age.